SentroProbe DR LS
The ATEX certified diffuse reflectance probes for SentroPAT FO systems offer powerful sampling options for solid dose pharmaceutical manufacturing processes in various unit operations. | |
|
- Two internal light sources
- Integrated white reference for outstanding long term stability
- 3/4" probe tip with different lengths
- GMP compliant design
- ATEX zone 20/21 certified
|
Parameter | Specification |
---|
Technology | Diffuse reflectance |
Light source | 2 redundant tungsten light sources |
Nominal life time | 14,000hrs for each bulb |
Lamp power | Configurable up to 5W |
Wavelength range | 400 - 2500nm |
Measurement spot | approximately 5mm |
Detection fiber | Single core 400µm fiber optical cable |
Measurement distance | Direct product contact |
Probe tip diameter | 3/4" (19mm) |
Standard lengths probe tip | 65mm, 100mm, 200mm, 300mm |
Ambient conditions | 0°C - 60°C, 0-10bar |
Installation | Flexible and welded 2" hygienic flanges |
Rating | IP65, ATEX zone 20/21, GMP compliant |